Electromagnetic NDT Acquisition System
InstrumentOperationalOpen for Testing / Collaboration

Electromagnetic NDT Acquisition System

Complete data acquisition setup for MBN, MIP and MNP measurements: function generator, AC power amplifier, preamplifier (x100), low-pass / band-pass filters, Hall sensor DC supply and NI-DAQmx acquisition.

Manufacturer

NI

Model

NI-DAQmx + BNC

Location

LEEC Laboratory

Technical Specifications

  • MBN, MIP, MNP equivalent sensing methods, BNC connector

Usage & Applications

  • Data acquisition for Magnetic Barkhausen Noise (MBN), Magnetic Incremental Permeability (MIP) and Magnetic Needle Probe (MNP) measurements.
  • Synchronized excitation (function generator + AC power amplifier) and acquisition (NI-DAQmx via BNC).
  • Processing of Barkhausen noise signals, MBNenergy envelopes and hysteresis cycles.
  • Structural health monitoring and material integrity evaluation campaigns.

Typical workflow:

1. Configure function generator excitation (H-field waveform). 2. Set preamplifier gain (x100) and filter (band-pass 1-90 kHz for MBN, low-pass 20 Hz for MIP). 3. Acquire signals through NI-DAQmx with BNC connector. 4. Post-process with signal analysis workstation (DeWeSoft) and export indicators.